Specifications 1-Year |
Fluke-721-1601 |
Range |
Resolution |
Accuracy |
Low pressure sensor |
-14 psi to +16 psi,
-0.97 bar to 1.1 bar |
0.001 psi,
0.0001 bar |
0.025% of full scale |
High pressure sensor |
-12 psi to +100 psi,
-0.83 bar to 6.9 bar |
0.01 psi,
0.0001 bar |
0.025% of full scale |
Fluke-721-1603 |
Low pressure sensor |
-14 psi to +16 psi,
-0.97 bar to 1.1 bar |
0.001 psi,
0.0001 bar |
0.025% of full scale |
High pressure sensor |
-12 psi to +300 psi,
-0.83 bar to 20 bar |
0.01 psi,
0.001 bar |
0.025% of full scale |
Fluke-721-1605 |
Low pressure sensor |
-14 psi to +16 psi,
-0.97 bar to 1.1 bar |
0.001 psi,
0.0001 bar |
0.025% of full scale |
High pressure sensor |
-12 psi to +500 psi,
-0.83 bar to 34.5 bar |
0.01 psi,
0.001 bar |
0.025% of full scale |
Fluke-721-1610 |
Low pressure sensor |
-14 psi to +16psi,
-0.97 bar to 1.1 bar |
0.001 psi,
0.0001 bar |
0.025% of full scale |
High pressure sensor |
0 psi to +1000 psi,
0.00 bar to 69 bar |
0.1 psi,
0.001 bar |
0.025% of full scale |
Fluke-721-1615 |
Range |
Resolution |
Accuracy |
Low pressure sensor |
-14 psi to +16 psi,
-0.97 bar to 1.1 bar |
0.001 psi,
0.0001 bar |
0.025% of full scale |
High pressure sensor |
0 psi to +1500 psi,
0.00 bar to 103.4 bar |
0.1 psi,
0.001 bar |
0.025% of full scale |
Fluke-721-1630 |
Low pressure sensor |
-14 psi to +16 psi,
-0.97 bar to 1.1 bar |
0.001 psi,
0.0001 bar |
0.025% of full scale |
High pressure sensor |
0 psi to +3000 psi,
0.00 bar to 200 bar |
0.1 psi,
0.01 bar |
0.025% of full scale |
Fluke-721-1650 |
Low pressure sensor |
-14 psi to +16 psi,
-0.97 bar to 1.1 bar |
0.001 psi,
0.0001 bar |
0.025% of full scale |
High pressure sensor |
0 psi to +5000 psi,
0.00 bar to 345 bar |
0.1 psi,
0.01 bar |
0.035% of full scale |
Fluke-721-3601 |
Range |
Resolution |
Accuracy |
Low pressure sensor |
-14 psi to +36 psi,
-0.97 bar to 2.48 bar |
0.001 psi,
0.0001 bar |
0.025% of full scale |
High pressure sensor |
-12 to +100 psi,
-0.83 bar to 6.9 bar |
.01 psi,
0.0001 bar |
0.025% of full scale |
Fluke-721-3603 |
Low pressure sensor |
-14 psi to +36 psi,
-0.97 bar to 2.48 bar |
0.001 psi,
0.0001 bar |
0.025% of full scale |
High pressure sensor |
-12 psi to +300 psi,
-0.83 bar to 20 bar |
0.01 psi,
0.001 bar |
0.025% of full scale |
Fluke-721-3605 |
Low pressure sensor |
-14 psi to +36 psi,
-0.97 bar to 2.48 bar |
0.001 psi,
0.0001 bar |
0.025% of full scale |
High pressure sensor |
-12 psi to +500 psi,
-0.83 bar to 34.5 bar |
0.01 psi,
0.001 bar |
0.025% of full scale |
Fluke-721-3610 |
Range |
Resolution |
Accuracy |
Low pressure sensor |
-14 psi to +36 psi,
-0.97 bar to 2.48 bar |
0.001 psi,
0.0001 bar |
0.025% of full scale |
High pressure sensor |
0 psi to +1000 psi,
0.00 bar to 69 bar |
0.1 psi,
0.001 bar |
0.025% of full scale |
Fluke-721-3615 |
Low pressure sensor |
-14 psi to +36 psi,
-0.97 bar to 2.48 bar |
0.001 psi,
0.0001 bar |
0.025% of full scale |
High pressure sensor |
0 psi to +1500 psi,
0.00 bar to 103.4 bar |
0.1 psi,
0.01 bar |
0.025% of full scale |
Fluke-721-3630 |
Low pressure sensor |
-14 psi to +36 psi,
-0.97 bar to 2.48 bar |
0.001 psi,
0.0001 bar |
0.025% of full scale |
High pressure sensor |
0 psi to +3000 psi,
0.00 bar to 200 bar |
0.1 psi,
0.01 bar |
0.025% of full scale |
Fluke-721-3650 |
Low pressure sensor |
-14 psi to +36 psi,
-0.97 bar to 2.48 bar |
0.001 psi,
0.0001 bar |
0.025% of full scale |
High pressure sensor |
0 psi to +5000 psi,
0.00 bar to 345 bar |
0.1 psi,
0.01 bar |
0.035% of full scale |
General Specifications |
Instrument setup recall |
Last used on power-up |
Environmental |
Operating temperature -10°C to +50°C
Storage temperature (without batteries) -20°C to +60°C |
Altitude |
2,000 m |
Ingress protection |
IP-40 |
Power requirements |
6.0 V DC |
Battery life |
>35 hours, typical usage |
Batteries required |
(4) Standard AA cells |
Dimensions (H x W x D) |
7.9 x 4.3 x 2.3 in (20 x 11 x 5.8 cm) |
Weight |
03539 kg (1 lb 3 oz) |
Connectors/ports pressure |
Two, 1/8 in NPT, 700P Series Pressure Module; RTD Probe |
Electrical and Temperature Measurement 1-Year |
Function |
Range |
Resolution |
Accuracy |
mA measure |
0 to 24 mA |
0.001 mA |
±0.015% of rdg
±0.002mA |
V DC measure |
0 to 30 V DC |
0.001 V |
±0.015% of rdg
±0.002 V |
Temperature measure (RTD/ Ohms)1 |
-40°C to 150°C
(-40°F to 302°F) |
0.01°C, 0.01°F |
±0.015% of rdg
±0.02 ohms
±0.1°C (±0.2°F)
±0.25°C (±0.45°F) combined uncertainty when using 720 RTD probe accessory |
Loop power supply |
24 V |
N/A |
24 mA @ 24 V |
1. Temperature measurement requires optional Fluke-720RTD Pt-100 RTD probe available as an accessory |
Temperature Effect |
No effect on accuracy on all functions from 15°C to 35°C |
Add ±0.002% F.S./°C for temps outside of 15°C to 35°C |
Available Pressure Engineering Units |
psi, bar, mbar, kPa, kgcm2, cmH2O @ 4°C, cmH2O @ 20°C, mH2O @ 4°C, mH2O @ 20°C, inH2O @ 4°C, inH2O @ 20°C, inH2O @ 60°F, mmHg @ 0°C, inHg @ 0°C, ftH2O @ 4°C, ftH2O @ 20°C, ftH2O @ 60°F |