7060 Really Cold Temperature Calibration Baths

These baths are completely self-contained. They require no auxiliary cooling fluids or devices to achieve their set-point temperatures. Using Fluke Calibration’s unique “heat-port” design, stability at –80 °C is ±0.0025 °C. No other company makes a bath that can match a Fluke Calibration bath’s performance, and Fluke Calibration baths are backed by our guarantee that if they don’t perform exactly the way we say they will, we’ll take them back. No arguments. No ifs, ands, or buts.

If you want to completely automate the entire calibration process, see the description of Fluke Calibration’s MET/TEMP II software package.

Forget commodity-like utility baths! They’re not designed for high performance calibration needs. And be careful of companies that advertise performance specifications they don’t meet. It’s easy to write down numbers; it’s more difficult to meet them with an instrument.

Remember, if our baths don’t perform the way we say they will, just send them back. Our equipment won’t disappoint you.

Chill to –40, –60, or –80 °C without external coolants

  • Self-contained refrigeration—no LN2 or chiller required
  • Temperatures as low as –80 °C in real metrology baths
  • Best stability and uniformity available at –80 °C and below
  • Large working areas for increased throughput

Find a refrigerated calibration bath for extremely low temperatures (i.e. –80 °C). Fluke Calibration metrology baths are well known world-wide for excellent temperature control that maintains excellent temperature, stability (±0.0015 °C) and uniformity (±0.003 °C). Models 7060 and 7080 can be used with Fluke Calibration bath fluids to reach extremely cold temperatures.

Do you need a bath that chills below –40 °C to temperatures as low as –60 °C or even –80 °C? Would you like a bath that reaches those temperatures without using any external coolants? Fluke Calibration the baths that meet these temperature requirements and give you the best stability in the industry.

Model Name Description
7060 Standard Bath, –60 °C to 110 °C
7080 Standard Bath, –80 °C to 110 °C

Accessories common to all models:

Accessory Description
Calibration Bath Fluids Full line of bath fluids
2010 Access Cover, 127 x 254 mm (5 x 10 in), Lexan
2007 Access Cover, 127 x 254 mm (5 x 10 in), Stainless Steel
2001-IEEE IEEE-488 Interface, 6330/7320/7340
2069 Magnifying Scope for Liquid-in-Glass Thermometer Calibration
2016-7080 Fluid Level Adapter, 7080
2016-7060 Fluid Level Adapter, 7060
2016-7060 Fluid Level Adapter, 7060
2620 RTD Probe, 280 mm (11 in.)

Specifications

7060

7080

Range −60 °C to 110 °C −80 °C to 110 °C
Stability ±0.0025 °C at −60 °C (methanol)
±0.002 °C at 0 °C (methanol)
±0.0015 °C at 25 °C (water)
±0.003 °C at 100 °C (oil 5012)
±0.0025 °C at −80 °C (methanol)
±0.0015 °C at 0 °C (methanol)
±0.0015 °C at 25 °C (water)
±0.003 °C at 100 °C (oil 5012)
Uniformity ±0.005 °C at −60 °C (methanol)
±0.005 °C at 0 °C (methanol)
±0.003 °C at 25 °C (water)
±0.005 °C at 100 °C (oil 5012)
±0.007 °C at −80 °C (methanol)
±0.005 °C at 0 °C (methanol)
±0.003 °C at 25 °C (water)
±0.005 °C at 100 °C (oil 5012)
Temperature Setting Digital display with push-button data entry
Set-Point Resolution 0.01 °C; high-resolution mode, 0.00007 °C
Display Resolution 0.01 °C
Digital Setting Accuracy ±1 °C
Digital Setting Repeatability ±0.01 °C
Heaters 500 and 1000 Watts
Access Opening 127 x 254 mm (5 x 10 in)
Depth 305 mm (12 in)
Wetted Parts 304 stainless steel
Power 230 VAC (±10 %), 50 or 60 Hz, 13 A, single phase, specify frequency
Volume 27 liters (7.2 gallons)
Weight 159 kg (350 lb.)
Size (HxWxD) 1168 x 775 x 483 mm (46 x 30.5 x 19 in)
Automation  Use the 1586A Super-DAQ to automate temperature sensor calibration