7040 Refrigerated Temperature Calibration Baths

These five baths operate at temperatures as low as –40 °C, and each one is built using CFC-free refrigerants. Fluke Calibration’s proprietary controller design and unique tank construction produce bath stabilities to ±0.001 °C or better. These baths are so stable and uniform that national labs use them for comparison calibrations and fixed-point cell maintenance.

Each bath (except the 7011) is fully automatable with a bath interface package and Fluke Calibration’s MET/TEMP II automation software package. When we automate a bath, we automate it completely with computer-controlled solenoid valves for precision balancing of the heating and cooling system. MET/TEMP II performs all calibration tasks automatically, using your PC.

With a Fluke Calibration cold bath, you can forget external coolants. Internal refrigeration systems are all that’s needed to reach each bath’s coldest temperature. Most cold baths may be ordered with an optional pumping lid for supplying external cooling requirements.

Each bath has unique characteristics that make it perfect for specific jobs. Some baths are excellent for SPRTs, some are great with thermistors, and some are perfect for maintaining triple point of water cells. A 7008IR bath can even be used to maintain the temperature of a blackbody cone.

Regardless of your application, Fluke Calibration has a bath that gets the job done, and done better than anyone else can do it. Call us today and tell us about your application.

Wide selection, with the best digital temperature control available

  • Stability to ±0.0007 °C
  • Best digital temperature controller available
  • “Super Tweak” function provides set-point resolution to 0.00003 °C
  • Excellent for maintaining fixed-point cells

Find a refrigerated calibration bath for low temperatures (i.e. –40 °C). Fluke Calibration metrology baths are well known world-wide for excellent temperature control that maintains excellent temperature stability (±0.0015 °C) and uniformity (±0.003 °C). Models 7008, 7040, 7037, 7011 and 7012 can be used with Fluke Calibration bath fluids to safely reach high and low temperatures.

Fluke Calibration’s temperature calibration baths are known around the world as the best calibration baths made. If you’re looking for a cold bath, no one gives you more choices than Fluke Calibration.

Model Name Description
7008 Standard Bath, –5 °C to 110 °C, high capacity
7011 Standard Bath, –10 °C to 110 °C
7012 Standard Bath, –10 °C to 110 °C, deep
7037 Standard Bath, –40 °C to 110 °C, deep
7040 Standard Bath, –40 °C to 110 °C
7008IR 7008 Bath, modified to accept an IR cone

Accessories common to all models:

Accessory Description
Calibration Bath Fluids Full line of bath fluids
2069 Magnifying Scope for Liquid-in-Glass Thermometer Calibration
2033 IR Cone (NIST design)
2071 Bath Cart, 7011, 7012 (312 mm [12.3 in] H)
2073 Bath Cart, 7008, 7037, 7040 (216 mm [8.5 in] H)
2011 Access Cover, 184 x 324 mm (7.25 x 12.75 in), Lexan (7008)
2010-5 Access Cover, 162 x 292 mm (6.38 x 11.5 in), Lexan (7037)
2027-5901 TPW Holding Fixture (7012, 7037)
2611 Thermistor Probe
2620 RTD Probe, 280 mm (11 in.)
2624 RTD Probe, 355 mm (14 in.)
Specifications
Model
7008
7040
7037
7012
7011
Range
–5 °C to 110 °C
–40 °C to 110 °C
–10 °C to 110 °C
Stability
±0.0007 °C at 25 °C (water)
±0.001 °C at 25 °C (mineral oil)
±0.002 °C at –40 °C (ethanol)
±0.0015 °C at 25 °C (water)
±0.003 °C at 100 °C (oil 5012)
±0.0008 °C at 0 °C (ethanol)
±0.0008 °C at 25 °C (water)
±0.003 °C at 100 °C (oil 5012)
Uniformity
±0.003 °C at 25 °C (water)
±0.004 °C at 25 °C (mineral oil)
±0.004 °C at –40 °C (ethanol)
±0.002 °C at 25 °C (water)
±0.004 °C at 100 °C (oil 5012)
±0.003 °C at 0 °C (ethanol)
±0.002 °C at 25 °C (water)
±0.004 °C at 100 °C (oil 5012)
Temperature Setting
Digital display with push-button data entry
Set-Point Resolution
0.002 °C; high-resolution mode, 0.00003 °C
0.01 °C; high-resolution mode, 0.00007 °C
0.002 °C; high-resolution mode, 0.00003 °C
Display Resolution
0.01 °C
Digital Setting Accuracy
±1 °C
Digital Setting Repeatability
±0.01 °C
±0.005 °C
Heaters
500 and 1000 Watts
Access Opening (call for customs)
324 x 184 mm(12.75 x 7.25 in)
127 x 254 mm(5 x 10 in)
162 x 292 mm(6.38 x 11.5 in)
127 x 254 mm(5 x 10 in)
Depth
331 mm (13 in)
305 mm (12 in)
457 mm (18 in)
305 mm (12 in)
Wetted Parts
304 Stainless Steel
Power
115 VAC (±10 %), 60 Hz, 14 A or 230 VAC, 50 or 60 Hz, 8 A, specify
115 VAC (±10 %), 60 Hz, 16 A or 230 VAC (±10 %), 50 or 60 Hz, 9 A (specify voltage and frequency)
115 VAC (±10 %), 60 Hz, 14 A or 230 VAC (±10 %), 50 Hz, 7 A, specify
Volume
42 liters (11.2 gal)
27 liters (7.2 gal)
42 liters (11.2 gal)
27 liters (7.2 gal)
Weight
61 kg (135 lb)
63.5 kg (40 lb)
68 kg (150 lb)
56.7 kg (125 lb)
Size
610 x 775 x 483 mm (24 x 30.5 x 19 in)
622 x 768 x 483 mm (24.5 x 30.25 x 19 in)
775 x 768 x 483 mm (30.5 x 30.25 x 19 in)
762 x 686 x 401 mm (30 x 27 x 15.8 in)
559 x 686 x 401 mm (22 x 27 x 15.8 in)
Automation Use the 1586A Super-DAQ to automate temperature sensor calibration