5080A High Compliance Multi-Product Calibrator

Accurate, reliable analog meter calibration

The 5080A Multi-Product Calibrator calibrates your analog workload accurately and reliably, thanks to its high voltage and current compliance. With maximum burden up to 800 mA for ac/dc voltage, and voltage up to 50 V for ac/dc current, 5080A calibrators can drive a wide range of analog meters.

Options and accessories expand workload coverage

Options and accessories enable you to use the 5080A Multi-Product Calibrator to calibrate an even broader workload, including:

  • Clamp meters. The 9100-200 10/50 turn coil and 5500A/COIL 50-turn current coil enables the 5080A to calibrate most popular clamp meters at currents up to 1000 A rms amps.
  • Oscilloscopes. Oscilloscope calibration to 200 MHz quickly, easily, and cost effectively. Verify dynamic response, bandwidth, timing, multiple triggering functions, and input resistance.
  • Megohm meters. This option sources high ohms, high voltage resistors up to 18 G ohms. It also measures high voltage outputs.

Protective circuits prevent damage

Mains voltage inadvertently applied to a calibrator’s output terminals can cause extensive damage, requiring costly repairs. Electrical protection for calibrator outputs is vital for daily operation. The 5080A calibrator’s innovative protection circuitry prevents it from being damaged by reversed input voltage, so you can use it with confidence day after day.

Highest voltage and current compliance

The 5080A has the highest voltage and current compliance of any calibrator in the Fluke Calibration multi-product and multifunction calibrator families, making it an ideal solution for calibrating analog meters and other instruments requiring higher drive capability for proper operation.

Maximum burden or compliance voltage
Functions
Model DC voltage AC voltage DC current AC current*
5080A 600 mA 800 mA 50 V 44 V
9100 20 mA 20 mA 4 V 4 V
5500A 10 mA 10 mA 4.5 V 3 V
5520A 10 mA 10 mA 7 V 5 V
5700A 50 mA 50 mA 10 V 7 V
5720A 50 mA 50 mA 10 V 7 V

* With AC LCOMP on

Automate the 5080A Multi-Product Calibrator to increase efficiency and throughput

For the full spectrum of calibration automation and asset management, choose MET/CAL software

The solutions you need, from the leader in calibration

Fluke Calibration is a premium brand, well known around the world for its accurate, dependable, high quality products. Long known as a leader in dc and low frequency ac calibration, Fluke Calibration is also recognized for its offerings in temperature, pressure, power, process, and rf calibration. Fluke Calibration provides the calibrators, standards, software, service, support and training you need for a complete solution in your cal lab.

Calibration solutions for your analog and digital workload

  • High compliance
  • Protection circuitry
  • Calibrates a wide workload, including analog meters and 3.5 and 4.5 digits DMMs

The 5080A Multi-Product Calibrator calibrates your analog and digital workload accurately and economically. Its high voltage and current compliance makes analog workload calibration easy and precise. And built-in protection circuitry protects it against damaging input voltages.

This easy-to-use multifunction calibrator can handle a wide meter calibration workload that includes:

  • Analog meters
  • Panel meters
  • Digital multimeters
  • Watt meters
  • Clamp meters (with coil accessory)
  • Megohm meters (optional)
  • Oscilloscopes to 200 MHz (optional)
  • …and more

Versatile software applications enable you to record paperless results, and more. Best of all, the 5080A Multi-Product Calibrator offers this versatile functionality at a price that fits nicely into your budget.

Model Name Description
5080A Multi-Product Calibrator
5080A/MEG Multi-Product calibrator with Megohm Meter Calibration Option
5080A/SC Multi-Product Calibrator with Oscilloscope Calibration Option
5080A/SC/MEG Multi-Product Calibrator with Megohm Meter and Oscilloscope Calibration Option

Accessories common to all models:

Accessory Description
9100-200 10/50 Turn Current Coil
5500A/COIL 50-Turn Current Coil
5080A/MEG 5080A Megohm Meter Calibration Option
5080A/CASE 5080A/CASE
5080A -> 5080A/SC/MEG Upgrade 5080A to 5080A/SC/MEG – Installable only at Fluke service centers for extra installation and calibration cost.
5080A -> 5080A/MEG Upgrade 5080A to 5080A/MEG – Installable only at Fluke service centers for extra installation and calibration cost.
Silver CarePlan Fluke Calibration Silver CarePlans are available for most calibration products. Please contact your local Fluke Calibration sales representative for details or to request a quote.
Priority Gold Instrument CarePlan Fluke Calibration Priority Gold Instrument CarePlans are available for most calibration products. Please contact your local Fluke Calibration sales representative for details or to request a quote.
Technical Specifications
Voltage DC
Range: 0 to ±1020V
Best 1 Year Specification: ±0.01% of setting
Current DC
Range: 0 to ±20.5A
Best 1 Year Specification: ±0.05% of setting
Resistance
Range: 0 to 190 MO, 19 values in x1 and x1.9
Best 1 Year Specification: ±0.025% of setting
Voltage AC
Range: 1 mV to 1020V
45Hz to 1 kHz, Sine
Best 1 Year Specification: ±0.1% of setting
Current AC
Range:29 mA to 20.5A
45 Hz to 1 kHz, Sine
Best 1 Year Specification: ±0.1% of setting
DC Power
Range:10.9 mW to 20.5 kW
Best 1 Year Specification: ±0.11% of setting
AC Power
Range: 10.9 mW to 20.5 kW
45-65 Hz, PF=1
Best 1 Year Specification: ±0.36% of setting
Phase
Range: 0 to ± 179.99°
45 Hz to 1 kHz
Best 1 Year Specification: ±0.25° of setting
Frequency
Range: 45 Hz to 1 kHz
Best 1 Year Specification: ±50 ppm of setting
General Specifications
Warm up Time
Twice the time since last warmed up, to a maximum of 30 minutes
Settling Time
Less than 5 seconds for all functions and ranges except as noted
Standard Interfaces
RS-232 and Ethernet
Temperature Performance
Operating: 0°C to 50°C
Calibration (tcal):15°C to 35°C
Storage:-20°C to +70°C
Temperature Coefficient
Temperature coefficient for temperatures outside tcal ±5 °C is 10 % of the stated specification per °C for temperatures in the range of 0 °C to 35 °C. Above 35 °C, the temperature coefficient is 20 % of the stated specification per °C.
Relative Humidity
Operating: <80% to 30°C, <70% to 40°C, <40% to 50°C
Storage:<95%, non-condensing
Altitude
Operating: 2,000m (6,500 ft) maximum
Non-operating: 12,200m (40,000 ft) maximum
Safety
Meets EN. 61010-1:2001, CAN/CSA-C22.2 No. 61010-1-04, UL 61010-1: 2004
Analog Low Isolation
20V
EMC
Meets EN613261:2006.
Line Power
Line Voltage (selectable): 100V, 120V, 220V, 240V
Line Frequency: 47 to 63 Hz
Line Voltage Variation: + 10% about line voltage setting
Power Consumption
600 VA
Calibration Documentation Report of calibration with data included; 17025 accredited report optional
Dimensions
Height:19.3 cm (7.6 in)
Width:43.2 cm (17 in), 44.3 cm (17.5 in) including handles
Depth:53.8 cm (21.2 in)
Weight (without options)
22 kg (48 lbs.)
Absolute Uncertainty Definition
The specifications include stability, temperature coefficient, linearity, line and load regulation, and the traceability of the external standards used for calibration. It is not necessary to add anything to determine the total specification for the temperature range indicated.
Specification Confidence Interval
99%